© 2009
Custom Sensor
Solutions, Inc.
List of Specifications for the M1010

Dilution range    2% to 100% (50-fold to 0-fold) set by front panel dial or remotely by analog voltage

Accuracy    2% of final concentration over 5% to 100% range

Flow rate    0 to 1000 cc/min, controlled by restriction of output port (can be used at higher flows with simple modification)

Inlet pressures    atmospheric only at both ports

Size    2.6" H x 8.2" W x 7" D

Power supply     12 VDC at 800 mA (supplied)

Materials contacting gas   

stainless steel, polypropylene, polyethylene, Viton® (in valve seal)
graphite composition (pumped version only)

Compatible gases (Note 1)   

all concentrations of air, oxygen, nitrogen, carbon dioxide, noncondensing water vapor, and inert gases

carbon monoxide (0.1 to 10,000 ppm)

nitrogen dioxide (dry only, 0.005 to 100 ppm)

nitric oxide (dryonly, 1 to 100 ppm)

hydrogen sulfide (dry only, 0.005 to 100 ppm)

sulfur dioxide (dry only, 1-500 ppm)

ethylene oxide (0.1 to 10,000 ppm)

most hydrocarbons and chlorohydrocarbons above 1 ppm

Incompatible gases (Note 1)   

hydrogen chloride, bromide, iodide, fluoride (all concentrations)

ammonia mixed with water vapor

combustible mixtures of all gases

benzene and other VOC's below 0.1 ppm


Note 1: The list of compatible and incompatible gases is not exclusive. The list of compatible gases includes only those that have been tested by us for extended periods.

Note 2: Viton® is a registered trademark of DuPont Dow Elastomers